Nubac Intradiscal Arthroplasty: Preclinical Studies and Preliminary Safety and Efficacy Evaluations

Qi-Bin Bao, PhD, Matthew Songer, MD, Luis Pimenta, MD, PhD, Dieter Werner, MD, Alejandro Reyes-Sanchez, MD, Massimo Balsano, MD, Umberto Agrillo, MD, Domagoj Coric, MD, Kenneth Davenport, MD, and Hansen Yuan, MD

MIS Society

BACKGROUND
Disc arthroplasty is gaining popularity for treatment of low-back pain caused by degenerative disc disease (DDD). It can involve total disc replacement or partial disc or nucleus replacement (or augmentation). Compared with total disc replacement, nucleus replacement is less invasive, has less surgical risk, has faster postoperative recovery, and doesn’t “burn bridges” should further surgery be required. However, nucleus replacement has a high risk of implant expulsion because the device is not fixed to the vertebrae. Nubac is the first polyetheretherketone (PEEK)-on-PEEK articulated disc arthroplasty device designed to optimally restore the lumbar anatomy and biomechanics.

METHODS
ISO 10993 standards were used to evaluate the biocompatibility of the PEEK material. Chemical and thermal–mechanical tests and in vivo study assessed PEEK’s biostability after exposure to high g irradiation and harsh oxidative conditions. Biomechanical tests to evaluate kinematic properties and anatomical restoration of the implanted lumbar motion segments and implant expulsion risk assessments were performed with a human cadaveric model. Because of the novelty of PEEK-on-PEEK as a self-mating articulating material, extensive wear tests were conducted with unidirectional and coupled motions. Static and fatigue strength also were tested. Animal study with a baboon model was conducted with gross, radiographic, biomechanical, and histological evaluations at 6 and 12 months postoperatively. Preliminary clinical data were collected through a prospective multicenter cohort study.

RESULTS
PEEK demonstrated exceptional biocompatibility and biodurability. Nubac restored disc height and motion segment range of motion. The unique articulating design of the Nubac demonstrated low risk of implant expulsion in a human cadaveric model. Wear tests showed that the Nubac has minimal wear and compares favorably to other disc arthroplasty materials. The Nubac also had excellent static and fatigue properties for the intended application. The animal study showed that the Nubac caused no adverse local or systematic tissue reaction and there was no detectable wear debris. The preliminary clinical data showed no major intraoperative vascular and neurological complications. There was significant Visual Analog Scale and Oswestry Disability Index score improvement.

CONCLUSIONS
The preclinical data supported the design rationale, and the preliminary clinical data (level II evidence) on safety and efficacy were encouraging.

CLINICAL RELEVANCE
The Nubac could be a viable first surgical option for patients with back pain caused by DDD.

Key Words disc arthroplasty, nucleus replacement, PEEK-on-PEEK articulation.
SAS Journal.
Winter 2007; 1; 36–45.
DOI: SASJ-2006-0007-RR

From the Department of Orthopedic Surgery, University of the Witwatersrand, Johannesburg, South Africa (Hähnle and Sweet); Linksfield Park Clinic, Johannesburg (Hähnle and Weinberg); Department of Medicine, University of the Witwatersrand (Sliwa); Department of Mechanical Engineering, University of Potchefstrom, Potchefstrom; South Africa (de Villiers); and Southern Medical, Wapadrand, South Africa (de Villiers).

Ulrich R Hähnle, Ian R Weinberg, and Malan de Villiers are codevelopers of the Kineflex disc and shareholders in SpinalMotion, Mountain View, Calif.

Address correspondence and reprint requests to Ulrich Hähnle, PO Box 52040, Saxonwold 2132, Johannesburg, South Africa (email: hahnleu@mdh-africa.org).

This submission was received January 8, 2006, and accepted for publication February 26, 2007.

We thank Stephanie Hanekom and Aldori Gehring for their enthusiastic help with patient management and follow-up.

The institutional review board of the hospital where all the cases have been performed (Linksfield Park Clinic, Nedcare, Johannesburg, South Africa) approved this study, which was conducted in accordance with the ethical standards of the Declaration of Helsinki (2000).

Top of Page

 

   
 
SAS    
The official publication of the Spine Arthroplasty Society SAS The official publication of the Spine Arthroplasty Society Published by RRY Publications - ryortho.com